23
Troubleshooting
Error Code
Severity
Possible Cause
Corrective Action
ASL - Please Close manual door
Warning
Stat door is open
Close the manual inlet /stat door
Autoslide Command Error
Warning
Autoslide command error
Retry the command
Autoslide Rejects Command Execution
Warning
The Autoslide is unable to execute the
command at this time
Action depends on the situation
Generally no action is required
Autoslide Weekly Shutdown Incomplete
Warning
An error occurred during the weekly shutdown
Check message log for other errors, correct problem
Reset Autoslide, and repeat weekly shutdown
Autoslide Weekly Startup Incomplete
Warning
An error occurred during the weekly startup
Check message log for other errors and correct problem
Reset Autoslide and perform weekly startup
Autoslide Weekly Startup Needed
Warning
Weekly startup is now required
Perform weekly startup
Note:
If the Autoslide is powered off, please wait 30
seconds before the power is switched back on.
This allows the ADVIA software to timeout
appropriately.