Troubleshooting
Error Code
Severity
Possible Cause
Corrective Action
ASL - Stainer error
Failure
Stainer error
Reset Autoslide
ASL - Stainer module door is open
Failure
Stainer module door is open
Close the strainer module door
Reset Autoslide / Reset Smearer
ASL - Stainer not primed
Failure
Stainer is not primed
Check message log to see if daily startup has been done
Perform a daily startup or Stainer Prime
(if daily startup was completed)
ASL - Stainer not ready
Failure
The stainer subsystem is not rinsed
Reset Autoslide
ASL - Stainer under methanol
Failure
Stainer wells are full with methanol
Run the daily or weekly startup procedure
ASL - Staining Error - Stop
Failure
Stainer door was open
Close stainer door
Reset Autoslide / Reset Smearer
ASL - Staining waste is full
ASL - Staining Waste Full - Stop
Failure
Staining waste is full
Dispose of staining waste according to local regulations
Reset Smearer
ASL - In Pre-smeared slide mode
Information
An operation is rejected because STAT
slide mode is in effect
Wait for Stat slide to be completed, close manual
slide entry door and try again
Note:
If the Autoslide is powered off, please wait 30
seconds before the power is switched back on.
This allows the ADVIA software to timeout
appropriately.