Troubleshooting
Troubleshooting
Error Code
Severity
Possible Cause
Corrective Action
ASL - Aspiration needle retract timeout
Failure
Aspiration needle failed to retract in time
Perform Reset Smearer
Resolve Autosampler error
ASL - Critical Error - Stop
Failure
General Autoslide critical error
Check and resolve errors on Message Log
Reset Smearer / Reset Autoslide
ASL - Dispenser error
Failure
Mechanical problem on slide dispenser
Inspect Slide Dispenser for damaged slide
Reset Smearer
ASL - Emergency stop cycle
Failure
Autoslide lower front cover opened during processing
Close lower front cover
Reset Smearer
ASL - Hardware error
Failure
A hardware error has occurred
Check message log for specfic errors and troubleshoot
ASL - Impossible to rinse
Failure
Unable to rinse sample or drop deposit lines; Other
command is processing
Reset Smearer
ASL - Operation completed in error
Failure
Run Stopped - Autoslide command failed
Check Message Log - Detail screen for specific cause
Resolve cause
Reset Autoslide
Note:
If the Autoslide is powered off, please wait 30
seconds before the power is switched back on.
This allows the ADVIA software to timeout
appropriately.