15
Monthly
Maintenance
Clean Stainer Wells and Plate after 3,000 Slides
1. Analyzer Mode: Standby
Autoslide Mode: Off
2. Slide open the stainer tray door. Manually
rotate the stainer plate and gently slide each
well out of the stainer plate.
3. Place the stainer wells in a tray filled with
methanol ensuring wells are completely
submerged.
4. Clean stainer plate surface with clean soft
cloth or paper towel.
5. Allow wells to soak for ten minutes, remove
from methanol and dry
with paper towel.
6. Replace stainer wells to stainer plate. Verify
wells are properly seated in stainer plate.
7. Close the stainer tray door.
8. Return analyzer to Ready mode.
Power on the Autoslide.