3D Reconstruction Upgrade Part 2
37
Page 37 of 46
Medical Solutions
Fig. 35:
Potentiometer
•
Move the C-arm into the -95° degree orbital position (limit switch = side of image inten-
sifier).
•
Loosen the Allen screw at the belt pulley
.
•
Attach the outside toothed belt (see
.
•
Use a screwdriver to turn the potentiometer counterclockwise into the end position
.
•
Once the potentiometer is in this position, turn it fully clockwise twice.
•
Secure the belt pulley with the Allen screw
.