Siemens AG
SPR2-320.814.02.01.02
ARCADIS Orbic
01.05
CS SD 24
3D Reconstruction Upgrade Part 1
25
Page 25 of 46
Medical Solutions
Fig. 17:
Connectors D200
NOTE
Cable ties or shielding can be removed as required during instal-
lation.
After successful installation, the shielding must be reestablished
and the cables secured with the cable ties.
•
(
) shows the layout of board D200.
•
Disconnect plug D1.X3 from PC board D1. Connect the plug to connector D200.X103
on interface PC board D200. Ensure proper locking.
•
On D30, reconnect plug D30.X5 to D200.X115 on interface PC board D200. For this
purpose, slip the cable out of the existing shielding and reinsert it from the opposite
side.
•
On D30, reconnect plug D30.X12 to D200.X112 on interface PC board D200. For this
purpose, slip the cable out of the existing shielding and reconnect it to the shielding at
D200 (interface)
. Retighten the shielding on PC board D30.
Cable no. 3, item number n.a.
0
•
Connect plug D30.X5 on D30 and plug-in location D200.X215 on D200.