4-
33D Reconstruction Upgrade Part 1
C-arm
0
•
Move the C-arm into the 0º position
.
•
Apply marks at 20 mm intervals to determine the lateral limit for the cam
.
•
Position the cam. Ensure that the cam is flush with the top and mark the holes with the
cam laterally at the 20 mm mark
.
•
Punch-mark the holes.
•
Ensure that shavings or dust do not drop into the C-arm.
•
Use a 3.2-mm bit for drilling. Continue until you have have drilled through the wall. En-
sure that you are not drilling beyond the wall
.
•
Counterbore the holes.
•
Cut the threads using an M4 thread tap (e.g. use a tap wrench)
.
•
Attach the cam using 2 countersunk screws M 4 x 8
.
•
Repeat these steps on the other side of the C-arm.
Fig. 3:
C-arm _ starting position