Testing the Protective Conductors
0
•
The system must be free of voltage.
•
Measuring de-
vice:
Protective conductor testing device; e.g., safety tester Unimed
1100
•
Measuring de-
vice:
Max. 0.2 Ohm
(observe country-specific regulations!) (current:
10 A)
•
Test procedure:
Measure between all touchable, conductive parts of the system
and the protective conductor wire at the ARCADIS Avantic
power plug.