Siemens AG
SPR2-330.815.01.01.02
ARCADIS Avantic
02.05
CS SD 24
General Startup
15
Page 15 of 46
Medical Solutions
Measuring the system leakage current
0
NOTE
Perform the system leakage current measurement only if the line
voltage on the ARCADIS Avantic system must be adapted to the lo-
cal line voltage.
See also
(Notes on equivalent leakage current
.
Measuring arrangements for system leakage current with tester VDE 0751 Octo-
ber 2001 (preferred measuring method)
0
Fig. 2:
Measuring circuit for measuring the system leakage current according to the differential
current method
•
Perform measurement configuration according to
. Application part
is not available in ARCADIS Avantic.
•
Perform measurement and repeat measurements.
•
Document the measured values in the "Leakage current test protocol",
.
Measuring arrangements for system leakage current with tester VDE 0751 Octo-
ber 1990 (e.g. Bender tester)
0