Specific System Settings
ARCADIS Avantic
SPR2-330.815.01.01.02
Siemens AG
02.05
CS SD 24
Page 28 of 46
Medical Solutions
Fig. 13: Buzzer setting
•
Under Configuration click
<Buzzer>
.
Make changes according to the following description:
•
Block time settings
-
Warning time limit:
The buzzer sound starts when the entered radiation time has
elapsed.
-
Block time limit:
After the entered radiation time has elapsed - without reset of the
"warning time limit" - the radiation is blocked the next time it is selected. Every reset
of "Warning Time Limit" also resets the "Block Time Limit" to "0".
•
Buzzer mode (indirect, I.I.)
-
Buzzer off:
Buzzer sound off. Exceptions: Push mode and "Warning Time Limit" has
ended.
-
Buzzer on:
Buzzer sound always during radiation.
The country-specific regulations must be observed for the max. fluoroscopy time and radi-
ation blockage.
Factory setting: