Parameter Function
Explanation
Q.EPS
Ratio correction
Ratio correction 80- 120 %
Q.FiL
Smoothing filter for min/
max memory
#
Off
no smoothing
time t
98
0-10 sec in 0.1 sec steps
Q.MEM
Min/max memory
Off
off
PhLd
Peak hold function
DTD
DTD function
Q.Phd
Hold time peak hold
function*
time t
98
in seconds (in steps of 0.1 sec)
Q.dtd
DTD limit**
See chapter 13.4
Q.Ano
Display mode during
sampling time
T
=
0
show lower limit of temp.
range during running
measurement
t.hld.
Hold previous temp.
reading during running
measurement
Q.dAP
Damping after min/max
memory*
Off
off
time t
98
0-999,9 sec in 0.1 sec steps
Q.drt
Pollution warning level
Value in % 0.1 - 100 %
Q.LiM
switch-off threshold
Value in % 0.1 - 100 %
Q.Si6
Q- signal intensity
Calculated signal intensity in %
Q.tMP
Q- temperature reading
Two colour/ratio temperature reading
End
End
*
#
The filtering affects the temperature reading and the signal intensity of
the two-colour mode.
*
* Only available with peak hold function
** Only available with DTD function