5
Operating instructions
5 Models
5.1 One-colour measurement
This measurement method is suitable for measurements of targets in areas where
no sighting obstructions, either solid or gaseous, exist. The target must fill the
entire target spot.
5.2 Two-colour measurement
The two-colour measuring method detects the infrared radiation at two different
wavelengths and calculates the quotient from the values obtained.
This measurement method is suitable for measurements of targets that are
partially obscured (either intermittently or permanently) by other objects, openings,
screens or viewing windows that reduce energy, and by dirt, smoke, or steam in
the atmosphere. The two color mode can also be used on targets that do not com-
pletely fill the target spot, provided the background is much cooler than the target.
Two-colour pyrometer
Type
Temp. range Focus
distance
Measuring
field size
LED spot
light
7MC3052-1AB68 550 - 1400 °C
1.5 m
Ø 21 mm
no
7MC3052-1AB66
700 - 1800 °C
0.2 m ... ∞
190 : 1
no
7MC3052-1AC66
1.08 m
Ø 5.6 mm
no
7MC3052-1AD66
0.12 m ... ∞
100 : 1
no
7MC3052-1AE66
0.033 - 0.045 m
50 : 1
no
7MC3052-1AF66
1.8 m
Ø 8 mm
no
7MC3052-1AF67 600 - 1400 °C
1.8 m
Ø 16 mm
no
7MC3052-1AB63
650 - 1600 °C
0.21 m
4.1 x 0.6 mm
yes
7MC3052-1AC63
1.0 m
18.5 x 2.7 mm
yes
7MC3052-3AB68
0.21 m
Ø 1.2 mm
yes
7MC3052-3AC68
1.0 m
Ø 5.6 mm
yes