User-defined calibration / scaling of the current out-
put
If necessary, the pyrometer can be adjusted with a user-defined calibra-
tion function. The following drawing explains the effects for offset and
factor.
Caution
:
To recalibrate your ARDOCELL PA, you will require a calibration
furnace and a reference standard.
In addition to a user-defined calibration, the user-defined range function
can also be activated. With this function being active, scaling of the cur-
rent output can be extended.
Please note:
T
shall
Current state
Influence - factor
Influence - offset
The measuring accuracy is not defined when the scal-
ing is outside the pyrometer range.
T
Pyrometer
Содержание 7MC3040-...
Страница 1: ...Digital Radiation Pyrometer PA 1x 2x 3x 7MC3040 Operating Manual 04 2017 ...
Страница 2: ......
Страница 8: ...s Operating manual PA ...
Страница 93: ...s Operating manual PA 85 29 Dimensions ...
Страница 97: ...s Operating manual PA 89 31 1 Mounting angle PA 11 K Pyrometer PA Mounting angle PA 11 K ...
Страница 98: ...s Operating manual PA 90 31 2 Polarising filter Pyrometer PA Polarising filter PA 20 P ...
Страница 99: ...s Operating manual PA 91 31 3 Mounting bracket PS 11 U Pyrometer PA Mounting bracket PA 11 U PA ...
Страница 101: ...s Operating manual PA 93 31 6 Cable VK 02 A Ident Nr 101 3909 ...
Страница 102: ...s Operating manual PA 94 31 7 Cable VK 02 F Ident Nr 103 1446 ...
Страница 103: ...s Operating manual PA 95 32 Mounting combinations 32 1 Mounting combination PA 20 007 ...
Страница 104: ...s Operating manual PA 96 32 2 Mounting combination PA 20 010 ...
Страница 111: ...s Operating manual PA 103 ...
Страница 112: ......
Страница 113: ......
Страница 114: ......
Страница 115: ......