Operation
5-30
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
5.9.4.1
Parameters: Pump/LCD Contrast: Adjust Pump
Here you can:
S
Increase
the pump capacity using the
→
or
↑
key
S
Decrease
the pump capacity using the
↓
key
S
Store the set pump capacity using the
ENTER
key
S
Cancel the input using the
ESC
key.
Changes to the pump capacity are shown on the flowmeter and directly
in the menu display by the message OK or NOK.
5.9.4.2
Parameters: Pump/LCD Contrast: Adjust LCD Contrast
Here you can:
S
Increase
the contrast using the
→
or
↑
key (darken the characters)
S
Decrease
the contrast using the
↓
key (brighten the characters)
S
Store the set contrast using the
ENTER
key
S
Cancel the input using the
ESC
key.
Notice
Press the three keys
↑
,
↓
and
→
simultaneously to set an average
contrast again.
Adjust pump/flow
Use
->
to adjust
ENTER: save new val.
ESC: keep old value
5941
Adjust LCD contrast
Use -> to adjust
ENTER: save new value
ESC : keep old value
5942
Содержание 7MB2335
Страница 2: ......
Страница 8: ...I 4 Operating Instructions ULTRAMAT 23 Gas Analyzer C79000 B5276 C216 03 ...
Страница 22: ...Installation Guidelines 2 8 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 112: ...Shut down 7 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 113: ...8 1 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Spare Parts List 8 ...
Страница 133: ...Spare parts list 8 21 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Страница 137: ...Spare parts list 8 25 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Страница 156: ...Appendix 9 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Your comments please ...
Страница 164: ...Appendix 9 12 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 166: ...Appendix 9 14 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 167: ......