
Technical Description
3-10
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
3.4 Integration
Communication
Communications facilities
The gas analyzers of series 6 (ULTRAMAT 6,
ULTRAMAT/OXYMAT 6, OXYMAT 6/61/64, FIDAMAT 6 and
CALOMAT 6/62), as well as the ULTRAMAT 23 offer the following
communications facilities:
• RS 485 interface
• PROFIBUS DP/PA
• AK interface (only OXYMAT 6, ULTRAMAT 6 and
ULTRAMAT/OXYMAT 6).
RS 485 interface (ELAN)
The standard integrated series interface permits to communi-
cate between several gas analyzers via the internal bus (ELAN).
Up to 12 gas analyzers with max. four components can be net-
worked.
The operating principle is shown in the following figure.
Typical design of an RS 485 network
.
Interface parameters
RS 485 interface (SIPROM GA)
SIPROM GA is a software tool for service and maintenance. All
analyzer functions (except factory functions) can be remote-
controlled and monitored via RS 485/RS 232 converters,
RS485/ethernet converters and RS 485/USB converters.
Up to 12 gas analyzers with max. four components can be net-
worked.
Networking of several gateways is possible when using the
RS 485/Ethernet converter (gateway). The number of operatable
analyzers is increased correspondingly.
Functions
• Display and saving of all analyzer data
• Remote operation of all analyzer functions
• Parameter and configuration settings
• Comprehensive diagnostics information
• Remote calibration
• Online help
• Cyclic saving of measured values and status on hard disk
• Exporting to commercially available application programs
• Downloading of new software.
•
Hardware requirements:
• PC/Laptop Pentium 133 MHz, RAM 32 MB, CD-ROM drive
• Free disk capacity min. 10 MB
• Free COM-Port: RS 232 or RS 485:
- for coupling to ELAN network RS 485/RS 232
- Linking the RS 485/Ethernet converter requires a 10 MB
standard network (RJ 45 connection) with TCP/IP. The dis-
tance should not exceed 500 m with an RS 485 network, a
repeater should be inserted for a longer distance.
Software requirements:
• Windows 95
• Windows 98
• Windows NT 4.0
• Windows 2000
• Windows XP.
Item
Designation
1
Computer
2
RS 485/RS 232 converter with RS 485/RS 232 cable
3
RS 485 bus connector with jumper
4
Analyzers
5
RS 485 cable
6
RS 485 bus connector
7
RS 485 network
8
9-pin SUB-D plug
9
Option: RS 485 repeater
Level
RS 485
Baud rate
9600
Data bits
8
Stop bit
1
Start bit
1
Parity
None
No echo mode
Ordering information
Order No.
Interface description
A5E00054148
RS 485/RS 232 converter
C79451-Z1589-U1
RS 485/USB converter
A5E00852382
Further accessories (e.g. cable,
connectors, repeater, etc.)
see Catalog IK PI
Содержание 7MB2335
Страница 2: ......
Страница 8: ...I 4 Operating Instructions ULTRAMAT 23 Gas Analyzer C79000 B5276 C216 03 ...
Страница 22: ...Installation Guidelines 2 8 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 112: ...Shut down 7 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 113: ...8 1 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Spare Parts List 8 ...
Страница 133: ...Spare parts list 8 21 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Страница 137: ...Spare parts list 8 25 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Страница 156: ...Appendix 9 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Your comments please ...
Страница 164: ...Appendix 9 12 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 166: ...Appendix 9 14 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 167: ......