Installation Guidelines
2-4
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
2.3
Gas Connections and Internal Gas Paths
A pipe with an outer diameter of 6 mm or 1/4” is present as the gas
connection.
Materials to be used: must be suitable for the respective measurement.
If the sample gas is to flow into a collective exhaust line, observe the
following points:
S
The exhaust line must be free of rapid changes in pressure. If this is
not possible, either use a separate exhaust line or
S
fit a damping vessel ( > 1 l) between the analyzer and the exhaust
line.
S
The exhaust line should always be routed with a falling gradient
since water make condense in it.
The corresponding gases must be sucked in via a fine filter. The
amount of the measured gas component must be negligibly small in the
AUTOCAL gas (zero gas). In particular when carrying out an
AUTOCAL for CO
2
ranges < 3 %, the air must be routed via a CO
2
absorber (e.g. soda lime).
With certain CO
2
ranges (see Section 3), the chopper section is purged
with clean nitrogen or CO
2-
free synthetic air at an inlet pressure of 300
to 350 kPa.
The internal atmospheric pressure sensor is routed to connection 6 in
the 19” rack version via a hose. It is therefore possible to connect the
pressure sensor (e.g. when using analyzer cabinets or houses) such
that it is guaranteed that only changes in atmospheric pressure are
recorded.
Refer to the Technical Description (Chapter 3) in this Manual.
Sample gas line
Paths for AUTO-
CAL/zero gas
Path for chopper
section purging
Path for pressure
sensor
Gas connections
and gas paths
Содержание 7MB2335
Страница 2: ......
Страница 8: ...I 4 Operating Instructions ULTRAMAT 23 Gas Analyzer C79000 B5276 C216 03 ...
Страница 22: ...Installation Guidelines 2 8 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 112: ...Shut down 7 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 113: ...8 1 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Spare Parts List 8 ...
Страница 133: ...Spare parts list 8 21 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Страница 137: ...Spare parts list 8 25 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Страница 156: ...Appendix 9 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Your comments please ...
Страница 164: ...Appendix 9 12 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 166: ...Appendix 9 14 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Страница 167: ......