Identifier: n =
0
= Measured
value
1
= Oven temperature ˚C
2
= Physical section temperature ˚C
3
= Pump temperature ˚C
4
= Pump temperature comparison ˚C
5
= Sampling temperature ˚C
6
= Sample gas probe temperature ˚C
7
= Flame temperature ˚C
8
= Temperature of flowmeter ˚C
9
= Temperature of board ˚C
10
= Flow
(ml/min)
11
= Pressure of sample gas/combustion
air (mbar abs.)
12
= Hydrogen pressure (mbar abs.)
13
= Fan current (mA)
14
=
FID voltage (V)
15
= +15
V
16
= –15
V
17
= +24
V
18/19/20/21/22/23
= –1/0/4/10/20/21
mA
24
=
raw current, corresponding to the
measuring range
25
=
current prior to zero, corresponding
to the measuring range
Display
CL
6
2
Identifier
*** 62 Reproduce internal values on analog signal,
channel 1
Test functions, RESET, switch-off
L
L
Test functions, RESET, switch-off
L
L
The analyzer attempts to ignite the flame for
approx. 20 minutes.
The ignition process starts automatically and
can be terminated by entering “0".
The procedure is terminated automatically when
the flame is ignited.
Factory setting: 0
*** 67 Ignite flame
Ignition on
Ignition off
1
6
0
CL
7
The sample gas pump is switched off in the
STANDBY status. Gentle backflashing of the
sample gas line takes place.
Factory setting: 1
*** 65 STANDBY
Measure
STANDBY
0
6
1
CL
5
This function must be used in the event of faults
in the program sequence. The analyzer is set to
the status of the warming-up phase.
*** 68 RESET
8
6
A-31