Further display functions
I
I
The current compensated in the zero is display-
ed here in pA.
15 Display compensated background current
(zero calibration)
Identifier: n =
1
= Oven temperature (100 to 200 ˚C)
2
= Physical section temperature
(10 to 60 ˚C)
3
= Pump temperature (110 to 190 ˚C)
4
=
Pump temperature comparison
(110 to 190 ˚C)
5
= Sampling
temperature
(110 to 190 ˚C)
6
= Sample gas probe temperature
(110 to 190 ˚C)
7
= Flame temperature (230 to 300 ˚C)
8
= Temperature of flowmeter
(10 to 80 ˚C)
9
= Temperature of board (10 to 70 ˚C)
10
= Flow (50 to 2000 ml/min according
to analyzer status)
11
= Pressure of sample gas/combustion
air (mbar abs.)
12
= Hydrogen pressure (
∼
1000 mbar
absolute)
13
=
Fan current (50 to 200 mA)
14
=
FID voltage (200 V)
15
= +15
V
16
= –15
V
17
= +24
V
Identifier: n =
0
= Measured
value
1
=
Raw measured value (pA)
2
= Analog
current
3
= Current following adjustment of
preamplifier
4
=
Current following zero adjustment
Terminating this function via CL key sets the
identifier automatically to 0.
Identifier n not
equal to 0
6
CL
n
1
Identifier
n
61 Display measured variables
6
CL
Display
0
Identifier
* 60 Display internal values
Display
5
1
A-27