System Overview
Troubleshooting
Troubleshooting Hardware
In addition to the corrective action listed for
common hardware troubleshooting, the following
is recommended:
• Do not submerse the epoc Host or
Reader into liquid
• Fold the Host and Reader together while
carrying to prevent dropping
• Do not remove the test card during sample
analysis, damage to the reader could occur
Hardware Troubleshooting
Scenario
Corrective Action
Host not charging
Ensure the Host blade is seated into
the Reader docking pivot
Host barcode reader
unable to scan
• Clean barcode reader with a dry
lint free cloth
• Perform a soft reboot by holding
the power button for 5 seconds
Host screen frozen
Perform a hard reboot by holding
the 9, 1 and power buttons
Reader red test indicator light on
View epoc Host screen for error
message and troubleshoot using the
epoc System Manual