System Overview
Additional Quality Assurance
Quality Control Fluid
Commercially available control fluids can be processed as determined by facility policy, federal, state and local requirements. It is recommended
to perform external quality control for each lot in each shipment of test cards. Refer to the epoc System Manual for recommended products.
Calibration Verification
Calibration verification is performed to verify accuracy of Test Results over an extended measurement range of a test. Performance of this
procedure may be required by regulatory or accreditation bodies. Refer to the epoc System Manual for recommended products.
Proficiency Testing
Follow Proficiency Testing procedure to verify the Accuracy and Precision of epoc System test results over multiple laboratories and/or sites.
Various laboratories can choose to register with different proficiency testing organizations.
Electronic
Value Assignment Datasheet (
e
VAD)
Download the current
electronic
V
alue
A
ssignment
D
atasheets from Siemens Document Library.
e
VADs are available for quality control and calibration verification fluids specific to the epoc system.
e
VADs are lot and software specific.