System Overview
Sample Processing
Sample Injection (Care-Fill™ Capillary Tube)
1. Once the sample has been collected, hold the capped tube by the
blue band and point the end of the capped tube upwards
2. Slide the adapter towards the end of the capped tube and insert
the tube into the adapter
3. Gently push the capped tube into the adapter end and gently push
until it clicks and locks in place
4. Hold the capped tube by the large end of the adapter and insert
into the sample entry port of the test card. Rotate the capillary
tube clockwise ¼ turn to ensure a good seal
5. While maintaining downward pressure, use the index finger of the
other hand to steadily depress the plunger with one continuous
motion until the Reader provides an audible beep and the Test
Status Indicator flashes green indicating enough sample was
received. The Host will display the message
Analyzing sample on
the Test Results page tab