Additional Quality Assurance
Performing Quality Assurance
1.
While connected to a Reader, select
Tools
2.
Select
Switch to QA test
or
From the epoc Readers screen tab, select and
hold the desired Reader
1.
Select
Run QA test
2.
Follow the instructions found in the sample
processing section for inserting a test card
3.
Scan the appropriate VAD or box barcode, or Enter
lot number manually and select the
icon to
search
4.
Using the Test type drop down menu, select the
desired QA test
5.
Use the Test Selection page tab to select the
desired analytes for the QA test. Depending on the
fluid type, all applicable analytes will be selected
automatically as soon as the barcode is scanned.
6.
Follow the instructions found in the sample
processing section for injecting a sample
7.
When the QA test is complete, the screen will
display Pass or Fail
10. Select the green bar to view the completed results
11. Remove the used test card and dispose in a biohazard receptacle
12. Insert a new test card and repeat previous steps if desired
13. When all testing with the Reader is complete, close testing by
selecting the Red X at the top right to close the Reader Screen
for that Reader.
Note:
Only authorized users will see the QA Test option available