5
Mounting
5.1
Mounting instructions
•
Observe the technical data.
•
Protect the sensor from direct sunlight.
•
To prevent condensation, avoid exposing the sensor to rapid changes in tempera‐
ture.
•
The mounting site has to be designed for the weight of the device.
•
To avoid inaccurate measurements when installing multiple devices: Make sure
that the laser light spot of one device is not in the visible range of another device.
•
Only commission the device 30 minutes after switching it on. Measured values
which are taken immediately after the device is switched on are not reliable.
5.2
Mounting device
1.
Mount the displacement measurement sensor using the designated fixing holes,
see "Setup and dimensions", page 12
2.
Make the electrical connection. Attach and tighten a voltage-free cable,
necting the device electrically", page 18
3.
Switch on the supply voltage.
✓
The green operating LED lights up.
The device needs around 10 seconds of initialization time before it is ready for
operation.
4.
Align the light spot so that the desired object is measured.
MOUNTING
5
8019642/ZJW1/2017-04-03 | SICK
O P E R A T I N G I N S T R U C T I O N S | OD1000
15
Subject to change without notice