Clearing Malfunctions
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
S045 Lamp energy
Lamp energy is outside tolerance range.
Check lamp energy in SOPAS
ET menu
Diagnosis/Sensor
values/Lamp
.
S047 High/low voltage
5 V or 24 V voltage outside tolerance range.
Please contact SICK Customer
Service
S048 Output no current
Output was switched to a currentless condition
because of time-out.
S049 Channel 1 error
The input range of the first analog input was
exceeded or the desired current at the first
analog input was not reached.
S050 Channel 2 error
The input range of the second analog input was
exceeded or the desired current at the second
analog input was not reached.
S051 Check sum error
The transmission process performed before-
hand from the master to the slave (controller)
has an incorrect check sum and the slave has
not accepted the data.
S052 Busy
The microcontroller of the module still performs
previous command.
S053 WARNxx
Internal warning.
Initiator Code Error text
Classifi-
cation
Description
Possible clearance
System S057 Communication problem Uncer-
tain
Communication fault between system and
subassembly.
Check cable connections in
system cabinet.
Restart the device.
S058 Configuration problem
Configuration error, found module does not
correspond to that of the nominal configuration
Please contact SICK Customer
Service
S059 Temperature
Temperature outside limit value.
Check temperature parameters
in menu 3.2
Diagnosis/Tem-
peratures
.
Check heating circuits.
S060 Watchdog OFF
Watchdog is disabled
Restart the device.
S061 FlashCard not detected
No flashcard
Please contact SICK Customer
Service
S062 Logbook problem
Logbook problem
Please contact SICK Customer
Service
Initiator Code Error text
Classifi-
cation
Description
Possible clearance
Initiator Code Error text
Classifi-
cation
Description
Possible clearance
System S065 Operational check
Opera-
tional
check
Operational check
---
Initiator Code Error text
Classifi-
cation
Description
Possible clearance
System S085 Module not found
Extended I/O (EXIST)
Please contact SICK Customer
Service
S091 Communication problem
Communication problem
S094 System start
System start
---