5.4.4.3
Test gas generator
Menu 2.4.3: Adjustment/Parameter/Test gas gener.
2.4.3
Measuring
Back
Save
Test gas gener.
/Adj/Para/Test
1 Span pt 1
2 Span pt 2
3 Span pt 3
4 Hg conc
3.0 µg/m3
6.3 µg/m3
9.0 µg/m3
9.3 µg/m3
The nominal values for 3 test gas concentrations of the
internal test gas generator CALSIC300 for automatic test
gas feed are entered in this menu.
Recommendation: 20%, 50%, 90% of full scale.
To change a value:
1
Select the desired span point.
2
“
Enter
”.
3
Change the value. To do so:
Enter the numeric value.
Minus sign: As first digit: Press
several times.
4
To save: “
Save
”.
5
The password is queried (
Measuring
Back
Save
Span pt 1
/Adj/Para/Conc/1
00003.000 µg/m3