5.4.4
Adjustment/Parameter
Menu 2.4: Adjustment/Parameter
Parameters for the checks are entered in these menus.
5.4.4.1
Span pt_manual
Menu 2.4.1: Adjustment/Parameter/Span pt manual
2.4
Measuring
Back
Enter
Parameter
/Adj/Par
1 Span pt manual
2 Correction factor
3 Test gas generator
4 Start time 1-8
5 Start time 9-16
2.4.1
The nominal values for 3 test gas concentrations of the
external test medium for manual test gas feed are entered
in this menu.
Recommendation: 20%, 50%, 90% of full scale.
Factor: Calculated correction factor
To change a value:
1
Select the desired span point.
2
“
Enter
”.
3
Change the value. To do so:
Enter the numeric value.
Minus sign: As first digit: Press
several times.
4
To save: “
Save
”.
5
The password is queried (
Measuring
Back
Enter
Measuring
Back
Save
Concentration
Span pt 1
/Adj/Para/Conc
/Adj/Para/Conc/1
1 Span pt 1
2 Span pt 2
3 Span pt 3
4 Factor
3.0 µg/m3
6.0 µg/m3
9.0 µg/m3
1.00000
00003.000 µg/m3