Operation
84
MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG
Su
bj
ect
to
ch
an
ge
with
ou
t n
otic
e
Settings
No.
1: Temperature of O
2
sensor
2: Temperature of detector
3: Temperature of cell
4: Temperature of detector unit
1
5: Temperature of heating controller 1, 1
st
controller, e.g. sample gas pump
6: Temperature of heating controller 1, 2
nd
controller, e.g. heating hose
7: Temperature of heating controller 2, 1
st
controller
8: Temperature of heating controller 2, 2
nd
controller
...
32: Temperature of heating controller 13, 2
nd
controller
33: Printer
The temperatures and the pressure can be interrogated in Specialist: System
edit: Results
(as of result 25) and in
Specialist: System edit: Zero/Calib/Purge/
other programs
(parameter TPA).
1
Only measurement, inputs are not processed
Name
Name of unit to be controlled. A selection list based on
Specialist: System edit:
Results
Actual
Actual value in °C resp. mbar (display, only)
Nominal
Nominal value in °C resp. mbar (can be entered)
Hyster.
Valid for temperature only.
The hysteresis (in °C) defines the switching thresholds of the temperature
controller
Actual > Hysteresis => Heating off
Actual < Nominal-Hysteresis => Heating on
Caution
: The first 3 heating controllers for O
2
sensor, Detector, Cell, cannot switch
a hysteresis. Set “0”.
Limit low
Temperature/pressure difference to lower limit in °C resp. mbar for error relay
Limit high
Temperature/pressure difference to upper limit in °C resp. mbar for error and
alarm relays
Offset
Temperature = ((Value of Controller) / Factor) + Offset
Value is fixed and cannot be changed.
Factor
Scaling factor Temperature = ((Value of Controller) / Factor) + Offset
Value is fixed and cannot be changed.
Relay ERROR
Relay number of error relay (inverted possible)
1: Actual > Limit high
1: Actual < Nominal-Limit low
0: Limit not exceeded
The relay is interrogated in Specialist: System edit: Zero/Calib/Purge programs/
other programs
.
Relay ALARM
Relay number of alarm relay (inverted possible)
1: Actual value > nominal value+limit high
Independent of a temperature decrease (external heating controller 5..32), the
relay remains set until RESET is carried out.
The relay is interrogated in Specialist: System edit: Zero/Calib/Purge/other
programs
. For RESET the error message ER 32
(
Specialist: System edit: Zero/Calib/Purge programs/other programs
is used.
Содержание MCS100E
Страница 1: ...O PERATING INSTRU CTIO NS MCS100E IR Analyzer Title Page Commissioning Operation Maintenance...
Страница 10: ...Contents 10 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG...
Страница 98: ...Operation 98 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...
Страница 112: ...Maintenance 112 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...