Maintenance
MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG
103
Su
bj
ect
to
ch
an
ge
with
ou
t n
otic
e
7.2
Visual check of the MCS100E
7.3
Check sample gas flow
1
View the “bar graph” display (F2 key).
2
View the “flow” bar graph (if necessary, press the “right” cursor key).
3
Compare the flow to the nominal value (application-dependent).
7.4
Check/replace the filter pad in the fan of the analyzer
The fan is located on the underside of the MCS100E.
It is not necessary to switch the MCS100E off to replace the filter pad.
1
Pull the cover off.
2
Take filter pad out.
3
Clean the filter pad or insert a new filter pad.
4
Press the cover on.
What is to be checked
Check
Ambient conditions
Installation location, ambient temperature, air humidity,
corrosive atmosphere.
Piping
Cracks, porous, brittle?
Enclosure and inside of device
Soiled, corroded? Loose parts? Odor? Unusual noise?
Are service and fault messages shown on
the display?
----
Spare parts required
Qty.
Part No.
Filter pad for analyzer
1
1
As required
1
5309683
Risk of soiling.
Operate the MCS100E without filter pad only for a short period.
Содержание MCS100E
Страница 1: ...O PERATING INSTRU CTIO NS MCS100E IR Analyzer Title Page Commissioning Operation Maintenance...
Страница 10: ...Contents 10 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG...
Страница 98: ...Operation 98 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...
Страница 112: ...Maintenance 112 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...