Operation
MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG
71
Su
bj
ect
to
ch
an
ge
with
ou
t n
otic
e
Calibration factor
This factor is used for multiplicative correction of the measured values.
It is calculated automatically upon calibration and can be corrected here, if
required. The calibration factor has an influence on the calculated
concentration.
Left column: Factor for span gas
Internal standard: Factor for internal standard
Extra Offset
Extra offset is used for additive correction of systematic drifts in addition to
the offset of the zero measurement.
Extra offset is not taken into account with zero measurement or calibration.
Extra offset is added to the calculated absorbance.
Extra factor
The extra factor is used for multiplicative correction of systematic derivations
from the calibration factor. The extra factor has an influence on the calculated
concentration.
The extra factor is not taken into account with zero measurement or
calibration.
Plausibility limits in % of
meas. range for Zero and
Calib. value
The value to be entered (default setting: 6%) defines the tolerance (in % of full
scale) which must not be exceeded by Zero resp. calib. gas sampling (since
the last monitoring).
If the value measured at Zero resp. calib. gas sampling
is within the tolerance
as entered here:
The measured value is accepted.
If the value measured at Zero resp. calib. gas sampling
exceeds the tolerance
as entered here:
The measured value is accepted, but a message is output to the printer and a
flag (Zero: ER17, Calib ER17: ER18) is set for interrogation in
Specialist:
System edit: Zero/Calib/Purge programs/other programs
.
0: No plausibility control,
calibration is always performed.
Oppress value
If the measured value is less than this value, then display THIS value.
First calibration value
Calibration value e.g. after service. Display in concentration.
Used to determine the drift.
Last calibration value
Calibration value of last calibration. Display in concentration.
Used to determine the drift.
Actual calibration value
Actual calibration value. Display in concentration.
First zero value
Zero value e.g. after service. Display in concentration.
Used to determine the drift.
Last zero value
Zero value of the last zero. Display in concentration.
Used to determine the drift.
Actual zero value
Actual zero value. Display in concentration.
Actual zero value [ABS]
Actual zero value. Display in absorbance.
Zero, CALIB all comp.
If a zero or calibration program is started with the option “all components”,
then THIS component is set (Yes) or not set (No).
Multiplication by result
Multiplication by the indicated result. This is used e.g. for printer correction.
0: No multiplication
Drift in [% measurement
range]
Drift in % of full scale
Drift total CALIB
Drift (span gas) since last drift reset (e.g. after service)
Drift last CALIB
Drift (span gas) since last drift correction
Factor
Equivalent to calibration factor (see above in this Table)
internal
Calibration factor for internal adjustment standard
Drift total ZERO
Drift (zero gas) since last drift reset (e.g. after service)
Drift last ZERO
Drift (zero gas) since last drift correction
Settings
Содержание MCS100E
Страница 1: ...O PERATING INSTRU CTIO NS MCS100E IR Analyzer Title Page Commissioning Operation Maintenance...
Страница 10: ...Contents 10 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG...
Страница 98: ...Operation 98 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...
Страница 112: ...Maintenance 112 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...