Monitor
Online help
LMS400 laser measurement sensor
5
Monitor
In the SOPAS ET monitor the measurement results from the LMS400 are displayed in
various forms. Use these displays to observe the measurement during operation or during
testing.
5.1
Scan view
Use the graphic scan view in SOPAS ET to verify the measured values generated and to
verify the measurement area online.
Notes
The scan view in the monitor is dependent on the available computing power of the
connected PC and is
not
output in real-time. For this reason not all measured values are
displayed. The same limitation also applies when saving measured values displayed in
a file.
The monitor displays the measured values
unfiltered
, i.e. the action of filters can
not
be checked with the aid of the monitor.
In the scan view you have the following additional options:
You can display distance values, energy values or distance and energy values.
You can change the diagram display.
You can save the scan data to a file.
How to display the measured values:
How to zoom in or zoom out the display:
Click on the Z
OOM
+ button.
The scan view is increased in size, the values on the diagram axes become smaller.
Click on the Z
OOM
– button.
The scan view is reduced in size, the values on the diagram axes increase.
How to display distance, power or both values:
Select the D
ISTANCE
option.
The distance data are displayed as a red line/dots.
Select the E
NERGY
option.
The energy data are displayed as a blue line/dots.
Select the D
ISTANCE
AND
ENERGY
option.
The distance data are displayed as a red line/dots, the energy data as a blue line/dots.
Open in the P
ROJECT
TREE
, LMS400-XX00, M
ONITOR
the device page S
CAN
VIEW
.
Click on the S
TART
button.
The scan line of the LMS400 is displayed.