Online help
LMS400
Configuration
8011984/YSZ3/2016-02-04
© SICK AG · Germany · All rights reserved · Subject to change without notice
15
Chapter
3
Invalid entries are displayed in SOPAS ET with a red border (see
).
Fig. 6:
Errors in the entry of the scan area
3.4.3
Other settings
In the M
ISCELLANEOUS
SETTINGS
area you can define a device name for the LMS400. The
device name will then be displayed on the project tree. Use this field, for example, to make
it easier to differentiate between different applications.
How to enter a device name:
If the PC is connected to the LMS400 and the device is switched online, the device name is
displayed in the project tree.
3.5
Advanced parameters
Note
The advanced parameters are only available in the “Authorised client” user level.
Enable extended RIS detectivity
If you want to measure objects with remission values smaller or higher than the values given
in the data sheet for the LMS400, you can extend the so-called Remission-Information-
System (RIS) in the LMS400. In this way the RIS facilitates, e.g., the improved detection of
dark or lighter bodies with reduced accuracy.
For measured values that are measured using extended RIS detectivity the following
applies:
The distance value may not have the defined measuring accuracy.
The same measurement point/the same object will not necessarily be detected by
another LMS400 because the remission value determined is dependent on the factory
calibration of the sensor, outside the specified range the calibration may be subject to
minor variations.
With extended RIS detectivity, your application must therefore evaluate, based on
additional criteria, whether the distance value is to be used or not.
Open in the P
ROJECT
TREE
, LMS400-XX00, P
ARAMETER
the device page B
ASIC
PARAMETER
.
In the M
ISCELLANEOUS
SETTINGS
area enter a device name with a maximum of 16
characters in the D
EVICE
NAME
field.
Press the Return key.
On this subject please read the section
in the chapter
“Product
description”
of the
LMS400 operating instructions .