7
M
aintenance and fault recti
fi
cation
7
.
1
M
aintenance
W
hen used as directed in normal operation
,
LBV
310
is completely
maintenance free
.
7
.
2
R
ectify malfunctions
T
he operator of the system is responsible for taking suitable measures
to rectify malfunctions
.
LBV
310
o
ff
ers maximum reliability
.
N
evertheless
,
faults can occur
during operation
.
T
hese may be caused by the following
,
e
.
g
.:
l
S
ensor
l
P
rocess
l
P
ower supply
l
S
ignal processing
T
he
fi
rst measure to be taken is to check the output signal
.
I
n many
cases
,
the causes can be determined this way and the faults recti
fi
ed
.
?
LBV
310
signals
"
covered
"
when the vibrating element is not
submerged
(
over
fi
ll protection
)
?
LBV
310
signals
"
uncovered
"
when the vibrating element is
submerged
(
dry run protection
)
l
O
perating voltage too low
à
C
heck operating voltage
l
E
lectronics defective
à
P
ress the mode switch
(
min
./
max
.).
I
f the instrument then
changes the mode
,
the instrument may be mechanically
damaged
.
S
hould the switching function in the correct mode
still be faulty
,
return the instrument for repair
.
à
P
ush the mode switch
.
I
f the instrument then does not change
the mode
,
the electronics module may be defective
.
E
xchange
the electronics module
.
à
C
heck if there is buildup on the vibrating element
,
and if so
,
remove it
.
l
U
nfavourable installation location
à
M
ount the instrument at a location in the vessel where no dead
zones or mounds can form
.
à
C
heck if the vibrating element is covered by buildup on the
socket
.
R
eaction when malfunc
-
tions occur
C
auses of malfunction
F
ault recti
fi
cation
C
hecking the switching
signal
LBV
310
• -
R
elay
(
DPDT
)
21
7
M
aintenance and fault recti
fi
cation
35919
-
EN
-
091126
Содержание LBV 310
Страница 1: ...D F GB S B E T R I E B S A N L E I T U N G LBV 310 Relais DPDT EN ...
Страница 31: ...ø 34 mm 1 34 178 mm 7 01 Abb 29 Temperaturzwischenstück LBV 310 Relais DPDT 31 9 Anhang 35919 DE 091125 ...
Страница 33: ...GB O P E R AT I N G I N S T R U C T I O N S LBV 310 Relay DPDT ...
Страница 62: ...ø 34 mm 1 34 178 mm 7 01 Fig 29 Temperature adapter 30 LBV 310 Relay DPDT 9 Supplement 35919 EN 091126 ...
Страница 63: ...LBV 310 Relay DPDT 31 9 Supplement 35919 EN 091126 ...