Substitute :
150°C
External :
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
No
back
Substitute :
150°C
External :
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
Yes
back
Substitute :
150°C
External :
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
Yes
back
Substitute :
150°C
External :
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
AnaIn
No
No
Figure 39: Activating the use of an external temperature sensor
9.4.7.3
Setting the temperature range: Lower limit value
Submenu item "Scale Low" serves to set the lower limit value for the sample gas
temperature. The lower limit value corresponds to Live Zero.
NOTE
Factory setting: 0 °C
Allowable values: See Technical Data
Contact SICK Service when you are not sure which limit value should be set.
1
Follow the menu path as described,
see "Parameter settings for the external
temperature sensor PT1000", page 59
.
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
back
Substitute :
150°C
External :
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
AnaIn
back
Substitute :
150°C
External :
Scale Low :
Scale High :
250 °C
Temperature Input
AnaIn
back
Substitute :
150°C
External :
Scale Low :
Scale High :
250 °C
Temperature Input
AnaIn
back
Substitute :
150°C
External :
Scale Low :
Scale High :
250 °C
Temperature Input
AnaIn
000 °C
000 °C
2 x
5 x
005 °C
back
Substitute :
150°C
External :
Scale Low :
5 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
AnaIn
Select
Select
Select
Figure 40: Setting the lower limit value for the temperature range (example) (corresponds to Live
Zero)
9.4.7.4
Setting the temperature range: Upper limit value (20 mA)
Submenu item "Scale High" serves to set the upper limit value for the sample gas
temperature.
NOTE
Factory setting: 250 °C
Allowable values: Max. 500 °C
Contact SICK Service when you are not sure which limit value should be set.
MENUS
9
8008932/16YT/V4-0/2022-02 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM901-02
61
Subject to change without notice