Figure 45: Entering the default value for the sample gas pressure (example)
9.4.10
Changing the Live Zero value of the analog output (Analog Out)
Submenu item "Analog Out" serves to change the Live Zero value.
NOTE
Factory setting: 4 mA
Min. value: 0 mA
Max. value: 4 mA
1
Follow the menu path as described,
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
Live Zero : 4 mA
back
Live Zero: 4 mA
edit: Enter
Analog Out
back
Analog Out
back
Analog Out
Select
Select
Parameters
back
back
Meas. Distance
Meas. Distance
Parameter Settings
back
back
Temperature
Temperature
Limit Value
Limit Value
Humidity
Humidity
Live Zero: 3 mA
back
Analog Out
Live Zero: 3 mA
edit: Enter
Figure 46: Changing the Live Zero value (example)
9.4.11
Changing the calibration values
Submenu item "Calibration" serves to change the characteristic gradient after a span
test and the offset correction for the zero point after a reference measurement during
on-site calibration.
1
Follow the menu path as described,
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
9
MENUS
64
O P E R A T I N G I N S T R U C T I O N S | GM901-02
8008932/16YT/V4-0/2022-02 | SICK
Subject to change without notice