TECHNICAL DATA
11
95
8027513/V1-0/2022-03|SICK
Subject to change without notice
O P E R A T I N G I N S T R U C T I O N S | MDA Vision
LED class
Risk group 1 (IEC 62471:2006-07, EN 62471:2008-09)
Temperature range
Operation: 0 °C ...+40 °C/Storage: -20 °C ... +50 °C
Relative humidity
0 … 90%, non
-condensing
Maximum ambient light
2,000 lx (on code)
Environmental
requirements
No vibrations, stable table needed (load capacity > 100 kg)
Supply voltage
SELV DC 24 V
Current consumption
max.
3.4 A
External power supply
AC 100 V ... 240 V, 50 Hz ... 60 Hz
Operational requirements
Object properties
Non-reflective, non-transparent
Object consistency
Dimensionally stable and detectable (all sides must be
visible)
No touching by operator during measurement
Surrounding conditions
Best measurement results are achievable if object is placed
in the center of the scale
Avoid direct sunlight on scale to improve results
Ensure a free space of min. 200 mm on both sides of the
scale (no objects placed) to avoid influences on measure-
ment
Uniform ambient light is needed for a good color image
quality
Remark
Foreign material (like stick out foil) may influence the
measurement
Intrusion during measurement may influence the
measurement
APPLICATION INSTRUCTIONS
To achieve the greatest possible measurement accuracy, the customer must ensure
the above requirements
Tab. 2: Data sheet of MDA Vision