No stable object weight
•
After the code has been read, the weight is displayed, but the value fluctuates.
•
As a result, dimension values can be determined, but no weight is displayed.
•
The weight value is > 0, but no stable weight could be determined within 5 seconds.
•
Vibration in the working environment.
•
Draft in the immediate vicinity.
▸
Remove the object from the scale.
This initiates a new measurement process.
▸
Place the object in the center of the scale.
Once the code is read, weight and dimension values are determined.
▸
NOTE
If a weight value still cannot be determined, make sure that the MDA Vision is set
up in a stable and vibration-free manner and that no possible drafts in the environment
affect the scale.
▸
If the problem cannot be solved by the customer, contact the service technician.
Error description
Fault cause