7.1.1   Measurements 

When the measured component is connected to the test leads, 
the screen displays information depending on the component 
and the selected settings in the R-L-C-D mode menu.  

        

 

 

 

 

Inductance 

 

Capacitance 

 

 

 

Resistance 

Diode 

 

In automatic mode, measurement of capacitance larger than 0.9 

μ

F by default is made at 120 Hz, while ESR is measured at 100 

kHz. In this case the display looks 
as follows. The frequency is not 
shown on the screen, but two 

additional numbers are displayed indicating the capacitor 
parameters described below.  

The main criterion for rejection of electrolytic capacitors is the 
value of ESR that varies amongst the types of capacitors and 
manufacturers. Two ESR tables have been compiled with typical 
ESR values for various capacitances/voltages, one for regular 
electrolytic capacitors and the other for Low ESR capacitors 
(presented below in Supplement A and B). These are indicative 
values and should only be used for a rough estimation of the 
capacitor performance. In this case besides the measured ESR 
value, two additional numbers are displayed in the right top 
portion of the screen. The first number indicates quality of the 
capacitor measured as a ln(Rs/Rt) where Rs is the measured ESR 
value and Rt is the value from the table. For standard aluminum 
capacitors a positive number indicate rejection, and for 

Low ESR

 

values above -10. The 2nd number is the selected operating 
voltage of the capacitor. It should be specified in the menu item 

Large Cap/C-voltage

. It should also be kept in mind that the 

measured capacitance may be significantly lower than the 
original data from the data sheet due to evaporation of the 
electrolyte and other degradation mechanisms. In such cases the 
table cannot be used for accurate capacitor evaluation. 

If you wish to measure capacitance larger than 0.9 

μ

F at a 

specific frequency you have to adjust frequency by pushing the 
joystick to the left for one beep or select frequency in the R-L-C-
D parameter sub-menu by pushing the joystick to the right for 
one beep to activate the hidden sub-menu. In this case both the 

Содержание LCR-Reader MPA

Страница 1: ......

Страница 2: ...Offset Calibration Board 11 7 2 R L C D mode menu 11 7 2 1 Primary Parameter 11 7 2 2 Secondary Parameter 12 7 2 3 Test Frequency 12 7 2 4 Test Signal Level 12 7 2 5 Period 12 7 2 6 SER PAR Mode 12 7...

Страница 3: ...13 4 4 Capacitance 23 13 4 5 DC Capacitance Measurement 24 13 4 6 Inductance 24 13 4 7 DC Voltage 24 13 4 8 AC Voltage 24 13 4 9 Low Frequency DDS Signal Generator 25 13 5 Supplement A 25 13 6 Supple...

Страница 4: ...ance Low ESR Ultralow Equivalent Series Resistance Rs Series Resistance Ls Inductance in series connection Cs Capacitance in Series Connection Xs Reactance in Series connection Rp Resistance in Parall...

Страница 5: ...sured component thus eliminating the offsets due to the device internal parasitics The measured component can be represented as one of the following equivalent circuits 1 and 2 AC measurement series c...

Страница 6: ...e same time the quantity of periods of the input signal is counted too Then the frequency f is calculated by the formula f M N fr where M is the number of periods of the input signal N is the number o...

Страница 7: ...ns and the number of beeps required to call the function The arrows show the direction of the joystick push A single arrow denotes holding for one beep two arrow heads denote holding for two beeps bef...

Страница 8: ...matically turn off at 3 5 V Information about the battery voltage can be displayed by selecting System Power Battery Voltage Charge the battery by connecting it via a micro USB cable to a USB DC volta...

Страница 9: ...ice mode and it is designated for measurement of Resistors Capacitors Inductances and Diodes To select the mode select R L C D in the main menu In order to get access to the mode parameters hidden sub...

Страница 10: ...only be used for a rough estimation of the capacitor performance In this case besides the measured ESR value two additional numbers are displayed in the right top portion of the screen The first numb...

Страница 11: ...value as well as the percentage deviation from the reference value To cancel Relative measurements push the joystick down for two beeps 7 1 4 Open Short Calibration Before the first use the device mus...

Страница 12: ...etween the test leads The dummy PCB uses holes to represent various sizes of components To use the calibration board place the test leads into the holes corresponding to the size of component under te...

Страница 13: ...used for measuring small values of capacitances and inductances as well as ESR Lower frequencies are used to measure large inductance and capacitance values Resistance measurements are always made at...

Страница 14: ...e and the values are held on the screen after the component has been disconnected from the device The device will only reset when a new component is measured or the test probes are shorted In order to...

Страница 15: ...and D Section 14 2 and 14 3 Two additional numbers are displayed at the top right corner of the screen in this case the first number is the quality of the capacitor For standard aluminum capacitors p...

Страница 16: ...at 1 3 V DC R test at 100 mV DC 7 3 1 Auto Mode To select the Auto Mode push the joystick down In the Auto mode resistances and diodes are automatically detected For resistors the bottom line displays...

Страница 17: ...cond line shows the peak to peak voltage range which is zero for a DC bias 7 4 1 Controls Push Joystick up for 1 beep to toggle the current voltage measurement mode In this mode the voltage value meas...

Страница 18: ...s mode it is possible to measure frequency pulse duration duty cycle period and count the number of pulses 7 6 1 Changing the sub mode Push the joystick to the right for 1 beep to open menu to select...

Страница 19: ...oes not turn off by flipping and the function of automatic change of the display orientation left right hand is disabled It is recommended to use LCR Reader Kelvin Probe Connector to simplify the devi...

Страница 20: ...e 1 s System Sound Medium All offsets are reset to zero 8 Device Troubleshooting 8 1 If the device does not turn on Push the joystick and hold it for 1 beep then release it Charge the battery Contact...

Страница 21: ...anty period the Manufacturer will at its discretion either repair or replace any product that proves to be defective To exercise this warranty write or call your local distributor You will be given pr...

Страница 22: ...arranties expressed or implied including any implied warranty of merchantability or fitness for a particular use The remedies provided herein are buyer s sole and exclusive remedies Neither Manufactur...

Страница 23: ...nd relative humidity up to 80 Dimensions Device 166x23x14 mm Case 180x40x20 mm Unit Weight 39 g 1 35 Oz Weight of the Device and Case 105 g 3 6 Oz 13 4 Details of the Measurement Modes 13 4 1 Test Sig...

Страница 24: ...0 1 0 2 3 10M 0 001M 0 5 5 20M 0 01M 3 5 Calibrate with short and open test leads before measurement 13 4 4 Capacitance Test signal 1 0 Vrms Range Resolution Test frequency 100Hz 120Hz 1 Hz 10 Hz 20...

Страница 25: ...0 H 0 1 H 1 3 0 2 3 0 5 2 1 2 10 mH 0 001mH 0 2 3 0 2 3 0 2 3 0 5 2 1 2 100mH 0 01mH 0 2 3 0 2 3 0 2 3 0 5 2 2 3 1000mH 0 1mH 0 2 3 1 0 3 0 3 3 3 5 10H 0 001H 2 3 2 2 2 3 100H 0 01H 5 3 5 3 Test signa...

Страница 26: ...Volts 6 3 10 16 25 35 50 63 100 160 1 0 142 118 74 2 2 64 54 34 3 3 43 36 22 4 7 30 25 16 6 8 21 17 11 10 14 12 7 4 22 6 5 5 4 3 4 33 5 4 3 3 6 2 2 47 4 3 5 3 0 2 5 1 7 68 3 5 2 7 2 4 2 1 1 7 1 1 100...

Страница 27: ...2 0 73 0 68 33 0 67 0 64 0 56 0 32 47 0 57 0 54 0 51 0 45 0 25 100 0 60 0 48 0 37 0 35 0 33 0 29 0 16 220 0 31 0 25 0 19 0 18 0 13 0 11 0 085 330 0 25 0 17 0 15 0 11 0 10 0 091 0 068 470 0 18 0 14 0 0...

Страница 28: ...te Quiet Exit Hand Brightness Contrast Exit Exit Save Mode Oscilloscope Decrease Parameter Increase Parameter T pix or V div Auto select T pix or V div Frequency Meter Period Frequency Counter Pulse D...

Отзывы: