48
-
When the Battery Voltage Has Dropped to 21 V or Less During Operation by BU-24B/BU-24L
(Sold Separately)
Power Supply
Fault Stop Setting
OFF Timer Selection
Manual Run
Time Stop
Time
Total
Volume
ON
(default setting)
The power
voltage fault is
displayed and
pump operation
is discontinued.
Data before the
stop will be
saved
.
The power
voltage fault is
displayed and
pump operation
is discontinued.
Data before the
stop will be
saved
.
The power
voltage fault is
displayed and
pump operation
is discontinued.
Data before the
stop will be
saved
.
The power
voltage fault is
displayed and
pump operation
is discontinued.
Data before the
stop will be
saved
.
OFF
Pump
operation is
continued even
if the battery
voltage drops.
There is the risk
that the
BU-24B/BU-24L
might
overdischarge.
So, set the
power supply
fault stop setting
to ON.
Even if the
battery voltage
drops, pump
operation is
continued until
the remaining
sampling time
reaches 0
(zero).
There is the risk
that the
BU-24B/BU-24L
might
overdischarge.
So, set the
power supply
fault stop setting
to ON.
Even if the
battery voltage
drops, pump
operation is
continued until
the sampling
stop time is
reached.
There is the risk
that the
BU-24B/BU-24L
might
overdischarge.
So, set the
power supply
fault stop setting
to ON.
Even if the
battery voltage
drops, pump
operation is
continued until
the total flow
rate reaches the
set value.
There is the risk
that the
BU-24B/BU-24L
might
overdischarge.
So, set the
power supply
fault stop setting
to ON.
7.9 Fault Stop Function
When a fault listed below occurs during sampling, the fault stop cause is displayed and sampling is
discontinued. Fault stop "enabling/disabling" can be set to "suction flow rate fault" and "power
voltage fault" in the fault stop settings. When a fault stop has occurred, data up to the point when
the fault stop occurred is recorded. Note, however, that when a power voltage fault has caused the
stop, all data may be lost depending on the timer setting.
List of Fault Stop Items
Displayed Message
Fault Stop Cause
Pressure Drop
Pressure loss fault:
A pressure loss of 40 kPa or more
continued for 30 seconds.
Ambient Temp.
Ambient temperature fault: An ambient temperature of 45
°
C or
higher continued for 30 seconds.
Flow Rate
Suction flow rate fault:
An error of
±
20 % or more of the set value
continued for one minute.
Power Low Vol.
Power voltage fault:
A power failure occurred or a
BU-24B/BU-24L (sold separately) voltage
of 21 V or less continued for 30 seconds.