5-4 Accumulation Time Setting
The accumulation time setting can be configured.
(No time setting configured)
→
Manual setting (default setting)
:
(Time setting configured)
→
Timer
setting
Press the [ENTER] key, which causes the hours display (hours) to flash. Then use the
[UP] or [DOWN] key to set the time (hours). Once again, press the [ENTER] key, which
causes the minutes display (minutes) to flash. Then use the [UP] or [DOWN] key to
change the setting for minutes.
Press the [ENTER] key to confirm the setting. (To return
to the previous screen without changing the setting, press the [MODE/ESC] key.)
Note:
The accumulation time can be configured between 1 minute and 9 hours 59 minutes.
5-5 Temperature Display Mode Setting
The temperature used to convert flowrates can also be set manually (default setting is
[AUTO]).
→
Automatic setting
(based on temperature sensor values)
°
C
→
Manual setting
(initially displays temperature sensor value)
It is recommended that this setting be used only if there is a temperature
sensor problem. The setting reverts to [AUTO] when the power is switched OFF.
No accumulation time
setting configured
(Hours setting)
(Minutes setting)
(Confirm manual setting)
Automatic setting
(numeric setting)
(Confirm)
Manual setting
(initially displays temperature sensor value)