24
Using an MP-W5P Mini Pump to Measure Low Flowrates (0.200 L/min or less)
Using an MP-
Σ
NII Series Mini Pump
The operating procedure is basically the same as for the MP-W5P, except for the
following points.
If the collection tube actually used is not available when using the FC-M1
flow calibrator to verify the MP-W5P mini pump flowrate, calibrate the
flowrate by connecting the low-flowrate orifice adapter included with the
product in front of the MP-W5P mini pump (as shown below).
Diameter adapter
Dedicated orifice
Dedicated communication cable
Diameter adapter
Vinyl tubing
4mm I.D.
×
7mm O.D.
×
60mm
Dedicated orifice
Low-flowrate
orifice adapter
●
Use an A-mini B type USB communication cable.
●
For MP-
Σ
30NII
mini pumps, use the ultra low flowrate suction holder
included with the pump to calibrate flowrates of 0.150 L/min or less.
●
For MP-
Σ
100HNII
mini pumps, apply an appropriate load during
calibration.