1
EXP. HOLD FUNCTIONS
104
EXP. HOLD FUNCTIONS
Chapter
1
16.
A: Valley hold
B: P-P hold
The minimum value and the difference between the maximum value and minimum value in
the section specified by the SECTION signal are held.
The value is maintained until the T/H signal turns ON.
17.
A: Average hold
B: Peak hold
The average of and the maximum value in the section specified by the SECTION signal are
held.
The value is maintained until the T/H signal turns ON.
When the Section setting is External
Sensor Input Value
Hold Value B
Hold Value A
Status Display
Status Display
Status Display
Status Display
ON
OFF
P-P
(
B
)
OFF
Blink
ON
OFF
ON
OFF
VLLY
(
A
)
OFF
Blink
ON
OFF
HOLD
(
A
)
OFF
HOLD
(
B
)
OFF
SECTION
T/H
H/E
Hold Value B
Hold Value A
Sensor Input Value
Status Display
Status Display
Status Display
Status Display
OFF
OFF
Blink
ON
OFF
PEAK
(
B
)
OFF
Blink
ON
AVE
(
A
)
HOLD
(
B
)
OFF
ON
OFF
HOLD
(
A
)
OFF
ON
OFF
H/E
T/H
SECTION