22
Troubleshooting
7 Troubleshooting
Precautions
WARNING
Troubleshooting should be performed only by personnel experi-
enced in RF systems and familiar with this equipment.
WARNING
The broadcast industry has recently recognized the potential med-
ical hazards of intense radio frequency radiation. Don't expose
personnel to personal harm. For reference, see CFR 29, Section
1910.97, the OSHA standard for exposure to non-ionizing radia-
tion. Whenever a rigger is on the tower in the area of the
antenna, shut off the transmitter and lock it off so that it cannot
be turned on accidentally.
CAUTION
Whenever you have the system open for repair, you must purge it
again as described in
Purging the System
on page 16. Never
begin operating the system under power until you are sure all the
moisture has been purged from it. You can do permanent damage
to the entire system, including the transmitter.
CAUTION
VSWR does not change of its own accord. If you find you must
repeatedly readjust the transformer to correct the VSWR, find and
correct the problem quickly. Otherwise, you will almost certainly
burn up your antenna and damage your transmitter. Look for the
cause in the following table.
Internal Arcing
Look for the cause of internal arcing in
Table 4
.
Table 4. Troubleshooting Internal Arcing
Possible Causes:
Cures:
Physical damage to transmission
line, feedline, or radiators.Dam-
age may have been caused by
ice, lightning, tower work, or
many other factors.
Damage may cause arcing
directly or by allowing water
inside the system.
Locate the damage. Replace dam-
aged components.
Purge the system after repair, in
accordance with
Purging the Sys-
tem
on page 16.
Missing or misaligned O-ring, if
the system has been opened re-
cently.
Locate the O-ring leak, using soap
solution. Replace the O-ring if
damaged.
Loss of pressurization.
Locate the leak. Re-purge in accor-
dance with
Purging the System
on
page 16 and restore pressurization.