background image

1. Model 

1.1 Model 

Model 

Measurement Range 

Output 

DSW-100-H 

5 to 90 %RH 

4 to 20 mA DC  2-wire type 

DSW-100-H W4 

4 to 20 mA DC  4-wire type 

DSW-100-HV W4 

0 to 1 V DC      4-wire type 

 

 

1.2 How to Read the Model Label

 

Model label is attached to the inside of the case.   
 

1

 Model 

2

 Measurement range, Output 

3

 Power supply voltage, Power consumption 

4

 Serial number 

 

(Fig. 1.2-1) 

 

2. Mounting 

 

Caution

 

Installation site should be examined, giving careful consideration to the following conditions.   

This sensor is designed for indoor use only. Do not install outside. 

[This instrument is intended to be used under the following environmental conditions.] 

• Free air flow 

• A minimum of dust, and an absence of corrosive gases 

• No flammable, explosive gases 

• No mechanical vibrations or shocks 

• No exposure to direct sunlight, an ambient temperature of 0 to 50   (32 to 122 ) that does not 

  change rapidly, and no icing 

• An ambient non-condensing humidity of 5 to 90 %RH 

• An absence of chlorinated and sulfidizing gases 

• No large capacity electromagnetic switches or cables through which large current is flowing 

• No water, oil, chemicals or the vapors of these substances can come into direct contact with the unit. 

 
 
 
 
 

Inquiries 

For any inquiries about this unit, please contact our agency or the vendor where you purchased the 

unit after checking the model and serial number. Please let us know the details of the malfunction, or   

discrepancy, and the operating conditions. 

 

 

SHINKO TECHNOS CO., LTD.

OVERSEAS DIVISION

:

Head Office

URL:

E-mail:

2-5-1, Senbahigashi, Minoo, Osaka, Japan

http://www.shinko-technos.co.jp/e/

[email protected]

Tel :

Fax:

+81-72-727-6100

+81-72-727-7006

4

3

2

1

Содержание DSW-100-H

Страница 1: ...lead to dangerous conditions and cause death or serious injury if not carried out properly Caution Procedures which may lead to dangerous conditions and cause superficial to medium injury or physical damage or may degrade or damage the product if not carried out properly Warning To prevent an electrical shock or fire only Shinko or other qualified service personnel may handle the inner assembly To...

Страница 2: ...sive gases No mechanical vibrations or shocks No exposure to direct sunlight an ambient temperature of 0 to 50 32 to 122 that does not change rapidly and no icing An ambient non condensing humidity of 5 to 90 RH An absence of chlorinated and sulfidizing gases No large capacity electromagnetic switches or cables through which large current is flowing No water oil chemicals or the vapors of these su...

Страница 3: ...ty sensitive so be sure to insert the sensor correctly otherwise the sensor will be broken and measurement will be impossible Never disassemble the humidity sensor as it may lead to malfunction 4 Mount the main unit cover then turn the power to the unit ON Socket Humidity sensor HD S2 3 Mounting Main Unit Screw 4 places C as indicated by the dotted lines and fix the sensor main unit to the Mountin...

Страница 4: ...m up period 60 minutes The output accuracy will be affected if the conditions above are changed Response characteristics Approx 20 seconds Time to reach 90 of the RH value when the RH shifts between 30 85 RH However airflow 5 l min 0 16 m s Waterproof filter attached Hysteresis Approx 0 RH Stabilization time 20 minutes Power Consumption Approx 550 mW 2 wire type Approx 3 W 4 wire type Operating En...

Отзывы: