![Shimadzu B252KS Скачать руководство пользователя страница 79](http://html1.mh-extra.com/html/shimadzu/b252ks/b252ks_instruction-manual_1256948079.webp)
-
76
-
4.4 Error
Display
List
4.4.1 General Display
Display Description
of
message
Wait for next display.
Minimum display resolution is decreased by one decimal
place.
Minimum display digit is returned to original state.
Date and time are being output.
Operation was aborted.
Application Measurement was released.
Calibration check detects too large error.
Calibration check detects too large error.
Menu lock is applied.
Menu lock is released.
Menu was reset.
Contents of new setting and coefficient were stored.
Reset by power failure.
Built-in weight is moving. Wait.
All numerals
blinking
Place the displayed calibration mass.
4.4.2 Error Display
Error display
Description
Countermeasure
The load on the pan is unstable at
calibration.
Only one of the calibration levers is
working.
Avoid wind and vibration.
The drift of zero point is large at
calibration.
Empty the pan.
The drift is large at the time of
PCAL.
Use
correct
weight.
The drift is large at span calibration.
Use correct weight.
Weight is wrong.
Use correct weight.
x
Stops when this is displayed.
*
Received command code is not
correct.
Check delimiter etc.
Integer of the displayed unit
exceeded 7 digits.
Decrease the load.
x
Failure in the balance.
*