76
-
2.4
General Menu Operations
2.4.1 Setting Numeric Values
Numeric values may be used to set the threshold of the comparator with the BX-K and
BW-K series balances.
Use the optional AKB-301 Application Keyboard to easily set these values. The values may
also be set using only the balance keys.
In a menu used to set numeric values,
and
are both illuminated and the digit to
be input blinks.
•
Press the
key to increase the value of the blinking digit by one.
( … →
)
•
Press the
key to move the blinking digit one place to the right.
•
Press the
key to store the displayed value in the balance memory.
is displayed when the value has been successfully saved.
is displayed when the balance failed to save the value.
•
Press the
key to stop numeric entry.
is displayed briefly and
the display returns to the menu, one level up.
2.4.2 Setting a Decimal Point
A decimal point is only used when setting units for solid density weighing, liquid density
weighing or when setting the multiplier for the user defined unit. Set the decimal point while
setting numerical values as follows.
•
Press the
key repeatedly until the last digit is blinking. Press the
key
once more to initiate decimal point setting mode.
The
▼
symbol or current decimal point blinks.
•
Press the
key to move the blinking decimal point one digit at a time to the
desired position.
•
Press the
key to set the decimal point position.
is displayed briefly to indicate that the setting is completed.
2.4.3 Menu Lock
The BX-K and BW-K series of balances have a "Menu Lock" function that locks the menu
selections to avoid changes being made by mistake.