Chapter 11 Troubleshooting
11-5
Alarm Code
35
Alarm Display
E-CAL Make Err.
Description
Creation of electronic test force calibration signal failed.
Corrective
Measures
The load cell must be unloaded when the E-CAL signal is created. Eliminate
any loads, then try again. Also, confirm whether the correct load cell capacity is
registered in the calibration cable.
Alarm Code
36
Alarm Display
E-CAL Err.
Description
Electronic calibration of test force failed.
Corrective
Measures
The load cell must be unloaded to perform E-CAL. Eliminate any loads, then try
again. Also, F-CAL must already be completed as well. If the error continues
after repeated attempts, it could be a hardware failure. Contact your Shimadzu
service representative.
Alarm Code
37
Alarm Display
Linearizing Err.
Description
Creation of nonlinear correction factor failed.
Corrective
Measures
Confirm whether the load direction (tension/compression) is correct. Also,
confirm whether the correct load cell capacity is registered in the calibration
cable. Confirm whether there was an error in the linearization procedure as
well.
Alarm Code
38
Alarm Display
Calibration Err.
Description
Calibration of analog input failed.
Corrective
Measures
If the alarm occurred during zero-adjustment
Confirm whether zero volts is applied to the analog input terminal.
If the alarm occurred during span-adjustment
Confirm whether 5 V or 10 V is applied to the analog input terminal.
Alarm Code
39
Alarm Display
E-CAL NonExe. ALM
Description
Attempted to start test without performing E-CAL after switching power ON or
after connecting calibration cable.
Corrective
Measures
Perform E-CAL.
Alarm Code
60
Alarm Display
Parameter Err. (1)
Description
The method used for the single test or cycle test is incorrect.
Corrective
Measures
Confirm the test speed and test direction.
Alarm Code
61
Alarm Display
Parameter Err. (2)
Description
The test method for test force control, strain control, or stroke control is
incorrect.
Corrective
Measures
Confirm automatic control parameters.
Alarm Code
63
Alarm Display
C.H. Return Err.
Description
Crosshead return was attempted without resetting position after changing test
direction or test force polarity.
Corrective
Measures
Reset the position whenever the test direction or test force polarity is changed.
Содержание AGS-100NX
Страница 2: ...This page is intentionally left blank...
Страница 16: ...Before Using the System xiv This page is intentionally left blank...
Страница 20: ...Contents xviii This page is intentionally left blank...
Страница 21: ...1 1 Chapter 1 Overview CONTENTS 1 1 Introduction 1 2 1 2 AGS X Unit Specifications 1 3...
Страница 27: ...2 1 Chapter 2 Part List Specifications CONTENTS 2 1 AGS X Unit 2 2 2 2 Peripheral Equipment 2 3...
Страница 38: ...Chapte 2 Part List Specifications 2 12 This page is intentionally left blank...
Страница 56: ...Chapter 3 Basic Instructions 3 18 This page is intentionally left blank...
Страница 118: ...Chapter 5 Performing Tensile Tests 5 12 This page is intentionally left blank...
Страница 126: ...Chapter 6 Performing Compression and Bending Tests 6 8 This page is intentionally left blank...
Страница 127: ...7 1 Chapter 7 Method Files CONTENTS 7 1 Introduction 7 2 7 2 File Operations 7 2 7 3 Computer File Operations 7 3...
Страница 130: ...Chapter 7 Method Files 7 4 This page is intentionally left blank...
Страница 131: ...8 1 Chapter 8 Control Input Output CONTENTS 8 1 Introduction 8 2 8 2 Registering Optional Devices 8 3...
Страница 134: ...Chapter 8 Control Input Output 8 4 This page is intentionally left blank...
Страница 142: ...Chapter 9 Sensor Input Output 9 8 This page is intentionally left blank...
Страница 146: ...Chapter 10 Maintenance and Inspection 10 4 This page is intentionally left blank...