High Accuracy Gloss Meter Operation Manual
9
4.2 Modify Calibration Parameters
Please modify the calibration plate parameters in the instrument simultaneously if
replace or re-calibrate the calibration tile.
Figure 15 Modify the calibration board parameters
Please enter the “calibration” menu to modify the calibration tile parameters, then
select the “modify parameters” menu item and press the OK button to enter the
“modify parameters” interface. Select the angle to be modified and press the OK
button to start editing the corresponding value. Press the measurement button to
save the changes after finishing the editing.
Note: The random change of the calibration tile parameters will lead to the
inaccuracy of measurement. Please consult the factory to re-calibrate the
calibration tile or replace the calibration box.
5. Measurement Modes
The instrument has four measurement modes. They are basic mode, quality control
mode, statistical mode and continuous mode.
Please do the simple measurement in the basic mode; the quality control mode can
compare the measurement result with the standard; the statistical mode can do
many average measurements, and perform basic statistics on multiple measured
data, also be compared with the standard; in the continuous mode, press the
measurement button to automatically measure continuously for the times you set.
5.1 Switching Measurement Modes
The instrument is in the basic mode when it is turned on for the first time. Please
enter the “main menu” and select the “measurement mode” menu item, then press
the OK button to the “Mode” interface to change the measurement mode as shown
in Figure 16. Select the appropriate measurement mode and press the OK button.