INSTALLATION 

Local city, county, or other ordinances may govern the use of this equipment.  If you have 
any  questions  about  local  requirements,  please  contact  the  appropriate  local  agency.  
Installation  may  be  performed  by  the  end  user.    It  is  unnecessary  for  this  unit  to  be 
installed by a technician. 

Under  normal  circumstances  this  unit  is  intended  for  use  indoors,  at  room  temperatures 
between  18

 and 28

C, at no greater than 80% Relative Humidity ( at 25

C ) and with a 

supply  voltage  that  does  not  vary  by  more  than  10%.    Customer  service  should  be 
contacted for operating conditions outside of these limits. 

This  unit  should  remain  upright  for  24  hours  prior  to  operating.    This  will  allow  the  oil  to 
settle in the compressor. 

2.1 

Power Source:

  See the incubator's serial/data plate for the voltage, cycle, phase 

and  ampere  requirements.  VOLTAGE  SHOULD  NOT  VARY  MORE  THAN  10% 
FROM  THE  DATA  PLATE  RATING.    These  units  are  intended  for  50/60  Hz 
application.  Electrical  supply  to  the  unit  must  conform  to  all  national  and  local 
electrical codes.  A separate circuit is recommended to avoid overloading or failure 
of other equipment on the same circuit. 

2.2 

Location:

    When  selecting  a  site  for  the  incubator,  consider  all  conditions  which 

may  affect 

performance

,  such  as  extreme  heat  from  steam  radiators,  stoves, 

ovens,  autoclaves, etc. Avoid direct sun, fast-moving air currents, heating/cooling 
ducts,  and  high  traffic  areas.  To  ensure  air  circulation  around  the  unit  allow  a 
minimum  of  20cm  between  incubator  and  any  walls  or  partitions  which  might 
obstruct free air flow. 

2.3 

Lifting  /  Handling:   

These  units  are  heavy  and  care  should  be  taken  to  use 

appropriate 

lifting

 devices that are sufficiently rated for these loads.  Units should 

only  be  lifted  from  their  bottom  surfaces.    Doors,  handles  and  knobs  are  not 
adequate for lifting or stabilization.  The unit should be completely restrained from 
tipping  during  lifting  or  transport.    All  moving  parts,  such  as  shelves  and  trays 
should  be  removed  and  doors  need  to  be  positively locked in the closed position 
during transfer to prevent shifting and damage. 

2.4 

Leveling:

    The  unit  must  sit  level  and  solidly.  Turn  the  leveling  feet 

counterclockwise to raise level. If the unit must be moved, turn the leveling feet in 
all the way to prevent bending or damage. 

2.5 

Cleaning:

    The  incubator's  interior  was  cleaned  at  the  factory, but not sterilizied. 

Remove all interior parts, including shelves and clean thoroughly with a disinfectant 
that  is  appropriate  for  your  application.    Regular  periodic  cleaning  is  required.  
Special  care  should  be  taken  when  cleaning  around  sensing  heads  to  prevent 
damage.    DO  NOT  USE  chlorine  based  bleaches  as  this  may  damage  the 
incubator  interior.    DO  NOT  USE  spray  cleaners  that  might  leak  through  cracks 
and openings and get on electrical components, or that may contain solvents that 
will harm coatings. 

WARNING:   

Never  clean  the  unit  with  alcohol  or  flammable  cleaners  with  the  unit 

connected to the electrical supply.  Always disconnect the unit from the electrical service 

Section 

Содержание RF02033111

Страница 1: ...LOW TEMPERATURE DIURNAL ILLUMINATION INCUBATOR MODEL RF2015 Serial Number RF02033111 INSTALLATION AND OPERATION MANUAL REV 09 10 4861428...

Страница 2: ...ROUBLESHOOTING SECTION 8 0 PARTS LIST UNIT SPECIFICATIONS WIRE DIAGRAMS This unit is a general purpose Diurnal Growth Chamber for professional industrial or educational use where the preparation or te...

Страница 3: ...r will arrange for official inspection to substantiate your claim 1 3 Return Shipment Save the shipping crate until you are sure all is well If for any reason you must return the unit first contact yo...

Страница 4: ...r circulation around the unit allow a minimum of 20cm between incubator and any walls or partitions which might obstruct free air flow 2 3 Lifting Handling These units are heavy and care should be tak...

Страница 5: ...when cleaning and assure all volatile or flammable cleaners are evaporated and dry before reattaching the unit to the power supply...

Страница 6: ...tes that you should consult your manual for further description and discussion of a control or user item 2 Indicates Temperature 3 Indicates Overtemperature 4 C Indicates Degrees Centigrade 5 Indicate...

Страница 7: ...ain Controllers set point If temperature rises to the Over Temperature set point this thermostat takes control of the heating element and allows continued use of the incubator until the problem can be...

Страница 8: ...ower switch to the ON position and turn each Overtemperature Thermostat to its maximum position clockwise using a coin or flat edged tool 5 2 Place a certified reference thermometer not supplied in th...

Страница 9: ...he day cycle first To stop the program push the EZ1 Button again Note When stopping the program the control will remain in the condition that it is previously in To restart the program push the EZ1 Bu...

Страница 10: ...EXT BOTTOM DISPLAY SCREEN PUSH UP OR DOWN ARROWS TO SELECT HOUR OF DAY IN UPPER DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN PUSH UP OR DOWN ARROWS TO SELECT MINUTES OF HOUR IN UPPER DISPLAY PUSH...

Страница 11: ...HANGE UPPER DISPLAY TO HOUR OF DAY PUSH ONCE FOR NEXT SCREEN USE UP OR DOWN ARROWS TO CHANGE UPPER DISPLAY TO MINUTE OF HOUR HOUR OF DAY MINUTE OF HOUR PUSH ONCE FOR NEXT SCREEN USE UP OR DOWN ARROWS...

Страница 12: ...H ONCE FOR NEXT BOTTOM PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN TO SELECT S tyP IN TOP DISPLAY USE UP OR DOWN ARROWS TO SELECT 12 IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN PUSH ONCE FOR...

Страница 13: ...IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT oFF IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT 4 IN TOP DISPLAY PU...

Страница 14: ...rent Set Point Temperatures This program uses one 1 file and six 6 different Steps Step 1 Is an unused Step and is reserved for starting the program at Specific Time Step 2 Parameters are for entering...

Страница 15: ...S TO SELECT DESIRED SET POINT IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN TO SELECT 0 IN TOP DISPLAY USE UP OR DOWN ARROWS TO SELECT 0 IN TOP DISPL...

Страница 16: ...OTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT on IN TOP DISPLAY USE UP OR DOWN ARROWS PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT 0 IN TOP DISPLAY hoUr hoUr 11 Min...

Страница 17: ...PLAY SCREEN Ent 1 USE UP OR DOWN ARROWS TO SELECT oFF IN TOP DISPLAY PUSH ONCE TO RETURN PREVIOUS SCREEN S tyP USE UP OR DOWN ARROWS TO SELECT SoAh IN TOP DISPLAY S tyP SoAh PUSH ONCE FOR NEXT BOTTOM...

Страница 18: ...FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT 2 IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DISPLAY SCREEN USE UP OR DOWN ARROWS TO SELECT 0 IN TOP DISPLAY PUSH ONCE FOR NEXT BOTTOM DIS...

Страница 19: ...N TO NORMAL DISPLAY A1 oPEr OR UNTIL SCREEN APPEARS AFTER THE OFFSET HAS BEEN ENTERED ALLOW UNIT TO STABALIZE RECHECK TEMPERATURE AGAIN AND IF CALIBRATION IS WITHIN EXCEPTABLE LIMITS THE OFFSET VALUE...

Страница 20: ...OR UNTIL SCREEN APPEARS nonE Usr S Usr S SET 1 PUSH ONCE PUSH TWICE 20 0 20 0 TO RETURN TO NORMAL DISPLAY SELECTING SET 1 UNDER Usr S SAVES THE OFFSET CALIBRATION FOR THE NIGHT CYCLE BEFORE CALIBRATIO...

Страница 21: ...in the chamber will spike and the Main temperature Controller will cycle off shutting down the heating element When the compressor is activated the temperature will stabilize at set point 6 2 Cleanin...

Страница 22: ...connection check shadow box for loose connections Unit will not heat over a temperature that is below set point 1 confirm that fan is moving and that amperage and voltage match data plate check fan mo...

Страница 23: ...ometer don t match 1 calibration error see section 5 9 2 controller failure evaluate if pilot light is operating correctly 3 allow at least two hours to stabilize 4 see if reference thermometer is cer...

Страница 24: ...not miss aligned 2 Steady internal clicking may be broken spring or valve call Customer Service MECHANICAL Motor doesn t move 1 if shaft spins freely check connections to motor and check voltage to m...

Страница 25: ...ce 2 check fuse circuit breaker on unit or in wall 3 see if unit is on e g fan or heater and just controller is off 4 check all wiring connections especially around the on off switch Contamination in...

Страница 26: ...t 4660505 4660505 High Limit Control 1750862 1750862 Interior Outlet 6100525 6100526 Interior Outlet European NA 6100531 Low Limit Control 1750538 1750538 Main Temperature Control 1750855 1750855 Pilo...

Страница 27: ...ons Shipping Weight 400 lbs Net Weight 280 lbs Exterior LxDxH in 34 5 x 32 x 77 Interior LxDxH in 27 x 23 5 x 57 Capacity 20 3 Cubic Ft Capacity 257 Bottles Temperature Range 10 to 40 C Temperature Un...

Страница 28: ...RED22G BLK22G SSRELAY 4 3 RTD AC 98 K2 W1 K3 99 L2 S1 R1 B5 D6 D5 L3 J3 DC Y1 NEUTRAL TERMINAL BLOCK BLU BLU RED RED YEL POWER SWITCH 9851057 dlr rev 6 22 09 POWERRELAY 115V 7030511 230V 7030520 7850...

Отзывы: