OPERATION 

 

P

UT THE 

I

NCUBATOR INTO 

O

PERATION

 

Putting the incubator into operation in a new workspace environment requires an 

8-hour period

 for 

the unit to come up to and stabilize at temperature prior to loading the incubation chamber with 
samples. During this period the incubator must be powered with the chamber door closed. Allowing 
time for stabilization helps protect samples. It is also necessary for the optional temperature display 
accuracy verification procedures, as well as any resulting calibration. 

Perform the following steps and procedures to prepare the incubator for use each time it is installed 
in a new workspace environment: 

 

1.

 

Plug the power cord into the workspace supply outlet 

2.

 

Place the Power switch in the ON ( I ) position 

3.

 

Perform the following Operation procedures in order: 

Optional: 

Set the Temperature Set Point

page 25. 

 

The incubator comes from the factory set to 20°C. 

Required

: Allow the incubator to thermally stabilize by operating sealed and 

undisturbed for 8 hours before continuing. 

Set the Over Temperature Limit

,

 page 26 

Load the Chamber

page 27 

 

End of procedure 

Содержание SRI3

Страница 1: ...BOD INCUBATOR 110 120 Voltage 220 240 Voltage Installation Operation Manual SRI3 SRI3 2...

Страница 2: ...er This unit is not designed for use with fruit flies Improper use of this unit including use with fruit flies will void any warranty Other units are specifically manufactured for fruit fly applicatio...

Страница 3: ...ollowing standards for use within an ambient air pressure range of 22 14 31 3 inHg 75 106 kPa with no flammable volatile or combustible materials being heated CAN CSA C22 2 No 61010 1 2012 CAN CSA C22...

Страница 4: ...djust Shelves 17 Deionized and Distilled Water 18 Installation Cleaning and Disinfection 18 GRAPHIC SYMBOLS 19 CONTROL PANEL OVERVIEW 21 OPERATION 23 Theory of Operation 23 Put the Incubator into Oper...

Страница 5: ...rator or their supervisors must draw up a site specific standard operating procedure SOP covering each application and associated safety guidelines This SOP must be written and available to all operat...

Страница 6: ...1 800 322 4897 extension 4 or 503 640 3000 FAX 503 640 1366 Sheldon Manufacturing INC P O Box 627 Cornelius OR 97113 ENGINEERING IMPROVEMENTS Sheldon Manufacturing continually improves all of its pro...

Страница 7: ...t chamber through the door space Select thermocouples suitable for the application temperature you will be calibrating at Why Probes Reference readings taken outside the chamber using wire temperature...

Страница 8: ...8 P a g e INTRODUCTION...

Страница 9: ...and exterior If you find any damage to the unit follow the carrier s procedure for claiming damage or loss 1 Carefully inspect the shipping carton for damage 2 Report any damage to the carrier servic...

Страница 10: ...10 P a g e Figure 1 SRI3 SRI3 2 RECEIVING YOUR UNIT ORIENTATION PHOTO Control Panel Temperature Probe Wire Shelf 1 of 2 Chamber Door Incubation Chamber Door Gasket...

Страница 11: ...The data plate contains the unit model number and serial number Tech Support will need this information during any support call Record it below for future reference The data plate is located near the...

Страница 12: ...12 P a g e RECEIVING YOUR UNIT...

Страница 13: ...supply is present page 15 Install the Incubator in a suitable location Review the lifting and handling instructions page 16 Install the incubator in its workspace location page 16 Make sure the incub...

Страница 14: ...units are built for use indoors at room temperatures between 15 C and 30 C 59 F and 86 F at no greater than 80 Relative Humidity at 25 C 77 F Operating outside these conditions may adversely affect th...

Страница 15: ...ge varies more than 10 The wall power source must be protective earth grounded Use a separate circuit to prevent loss of the unit due to overloading or circuit failure The wall power source must confo...

Страница 16: ...uate for lifting or stabilization Restrain the unit completely while lifting or transporting so it cannot tip Remove all moving parts such as shelves and trays and lock doors in the closed position du...

Страница 17: ...is used to secure the shelves during transit and prevent damage to the chamber interior 1 Remove all shipping materials from the shelving Optional Shelf Adjustment Move the shelves as needed for your...

Страница 18: ...0 uS cm for cleaning applications INSTALLATION CLEANING AND DISINFECTION The manufacturer recommends cleaning the shelving and unit chamber prior to installation of the shelving in the chamber The un...

Страница 19: ...ge de la temp rature Over Temperature Limit system Thermostat temp rature limite contr le haute AC Power Rep re le courant alternatif I ON O OFF I indique que l interrupteur est en position marche O i...

Страница 20: ...20 P a g e GRAPHIC SYMBOLS...

Страница 21: ...rrent incubator air temperature accurate to 0 1 C The Up and Down buttons are used to navigate to the temperature set point and calibration adjustment display modes The display blinks continually whil...

Страница 22: ...22 P a g e CONTROL PANEL OVERVIEW...

Страница 23: ...to avoid overshooting A circulation fan provides even air distribution throughout the chamber and plays an important role in maintaining temperature uniformity around the shelf space Temperature Low...

Страница 24: ...uracy verification procedures as well as any resulting calibration Perform the following steps and procedures to prepare the incubator for use each time it is installed in a new workspace environment...

Страница 25: ...ment mode after 5 seconds of inactivity with the last shown set point value saved Set Point Adjustment Mode Initial Set Point 3 Adjust the set point to your application temperature using the Up or the...

Страница 26: ...he unit temperature stability Check if any powered accessories in the incubation chamber are generating heat If you find no sources of external or internal temperature fluctuations contact Technical S...

Страница 27: ...BILITY AND POWER OUTLET Outlets The SRI3 is provided with 2 power outlets inside the chamber Do not attach powered equipment drawing more than 1 amp from both outlets The SRI3 2 is provided with 1 pow...

Страница 28: ...oint up and down in small increments If the relative humidity in the incubation chamber is already near the dew point barometric fluctuations may push it across the dew point threshold Note Thin air a...

Страница 29: ...less steel surfaces Do not use chlorine based bleaches or abrasives these will damage the chamber liner Warning Exercise caution if cleaning the unit with alcohol or flammable cleaners Always allow th...

Страница 30: ...an inhibit growth or cause metabolic symptoms in microbiological sample populations Make sure that chlorines quaternary ammonias or any other overtly volatile disinfecting agents have been rinsed or o...

Страница 31: ...below freezing This happens frequently when the unit is run toward the low end of its performance range At these temperatures humidity in the air will condense and freeze on the coil Figure 3 Refrige...

Страница 32: ...ates it insulates the heat transporting coolant fluid within the coil driving an increasing loss of chilling efficiency This may even cause false high temperature readings on the defrost thermostat an...

Страница 33: ...ith the door open for 24 hours 2 Wipe up any condensate that forms in the chamber as well as any water around the incubator after 24 hours have elapsed 3 Close the chamber door plug the incubator powe...

Страница 34: ...nd use the calibration setup required by your industry requirements or laboratory protocol Suggested Single Point Offset Calibration Setup Non stick tape recommended 1 Introduce the reference device t...

Страница 35: ...ture Calibration 1 Once the incubator temperature has stabilized compare the reference device and incubator temperature display readings If the readings are the same or the difference between the two...

Страница 36: ...ber temperature value 4 Use the Up and Down buttons to adjust the current display temperature value until it matches the reference device temperature reading Reference Device 5 After matching the disp...

Страница 37: ...l range Reference Device 8 If the two readings are not the same and the difference still falls outside the acceptable range of your protocol repeat steps 3 7 up to two more times Three calibration att...

Страница 38: ...38 P a g e USER MAINTENANCE...

Страница 39: ...f the height width and depth of the inner chamber All indications are average values typical for units produced in the series We reserve the right to alter technical specifications at all times WEIGHT...

Страница 40: ...SRI3 10 0 lb 4 5 kg 20 0 lb 9 0 kg SRI3 2 10 0 lb 4 5 kg 20 0 lb 9 0 kg TEMPERATURE Model Range Uniformity Stability SRI3 Ambient 0 C to 45 C 0 5 0 2 C SRI3 2 Ambient 0 C to 45 C 0 5 0 2 C POWER Mode...

Страница 41: ...322 4897 extension 3 If you are uncertain that you have the correct Part Number or if you need that specific item please contact Sheldon Technical Support for help at 1 800 322 4897 extension 4 or 503...

Страница 42: ...P O Box 627 Cornelius OR 97113 USA support sheldonmfg com sheldonmanufacturing com 1 800 322 4897 503 640 3000 FAX 503 640 1366...

Отзывы: