OPERATION
S
ET THE
T
EMPERATURE
S
ET
P
OINT
The incubator comes from the factory with a set point of 20
°
C. Perform the steps below to change
the set point to your process or application temperature.
End of Procedure
Set Temperature Set Point
1
Turn the
OTL
dial clockwise to its max setting, if not already
set at max.
•
This prevents the heating cutoff system from interfering
with this procedure.
2
Briefly push and release either of the temperature arrow
buttons to activate the temperature set point adjustment
mode.
•
The display will briefly flash the letters “SP”, then show
the flashing, adjustable temperature set point.
Note:
The display will automatically exit the adjustment mode
after 5 seconds of inactivity, with the last shown set point
value saved.
Set Point
Adjustment Mode
Initial Set Point
3
Adjust the set point to your application temperature using the
Up
or the
Down
arrow keys.
New Set Point
4
Wait 5 seconds after entering the set point.
•
The display will stop flashing, and the set point is
now saved in the controller.
•
The chamber will now automatically heat or cool to
match your set point.
•
The display will revert to showing the current
chamber air temperature.
Heating to New Set
Point
5
See the
on page 26 for how to set the OTL
system once the incubation chamber has stabilized at your application
temperature set point and after you have performed any display
verifications or calibrations.
Note
:
When putting the unit into operation in a new location, allow the
incubator to run 8 hours prior to setting the OTL.
Содержание SRI3
Страница 1: ...BOD INCUBATOR 110 120 Voltage 220 240 Voltage Installation Operation Manual SRI3 SRI3 2...
Страница 8: ...8 P a g e INTRODUCTION...
Страница 12: ...12 P a g e RECEIVING YOUR UNIT...
Страница 20: ...20 P a g e GRAPHIC SYMBOLS...
Страница 22: ...22 P a g e CONTROL PANEL OVERVIEW...
Страница 38: ...38 P a g e USER MAINTENANCE...